5. SENSORS

The following Pt 100 Ω sensors can be used with the C.A 865
thermometer.
Model
Reference Measurement Response
SP10 surface P03652712 -50...+200°C
Sensitive element is spring-mounted to ensure optimal contact even if
the sensor is not placed at right angles.
SP11 needle
P03652713 -100...+600°C
For penetration of viscous, thick substances
SP12 air
P03652714 -100...+600°C
Suited to all ambient measurements in air.
SP13 liquid
P03652715 -100...+600°C
For penetration of viscous, thick substances

6. MAINTENANCE

!
For maintenance, use only specified spare parts. The
manufacturer will not be held responsible for any accident
occuring following a repair done other than by its After Sales
Service or approved repairers.
6.1
Maintenance
6.1.1
Battery replacement
I Press on the ON/OFF pushbutton
!
Take off the cover on the back of the instrument
I
Replace the worn-out battery with a 9 V battery (type 6LR61
I
or 6LF22)
6.2

Cleaning the casing

Clean the casing using a cloth lightly dampened in soapy water.
Rinse off with a damp cloth.
Do not use solvent.
!
6.3
Metrological verification
Like all measuring or testing devices, the instrument must
!
be checked regularly.
This instrument should be checked at least once a year. For
checking and calibration, contact one of our accredited
metrology laboratories (information and contact details available
on request), at our Chauvin Arnoux subsidiary or the branch in
your country.
Dimensions
range
time
(mm)
6 s
Ø 5 - L = 130
7 s
Ø 3 - L = 130
5 s
Ø 5 - L = 130
7 s
Ø 3 - L = 130