Start or stop measurements on a channel (2- or 4-channel analyzers)

1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > CONFIGURE
SEQUENCER > ACTIVATE CHANNELS.
2. Select a channel to start measurements. Unselect a channel to stop measurements. Push the
LEFT arrow to select or unselect a checkbox.

Change the channel measurement order (2- or 4-channel analyzers)

To change the order that the channels are measured, do the steps that follow:
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > CONFIGURE
SEQUENCER > SEQUENCE CHANNELS.
2. Push the UP and DOWN arrows to select a row.
Note: S1 is the first channel measured, followed by S2, S3 and S4.
3. Push the LEFT or RIGHT arrow to select a channel.
Note: Do not select channels that contain the symbol "~" (e.g., 4-~SAMPLE4). Channels that contain the
symbol "~" are not measured.

Set the date and time

Set the date and time format and the date and time that show on the measurement screen and in the
log files.
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > SET DATE/TIME.
2. Select an option.
Option
Description
DATE FORMAT Sets the date format (YYYY= year, MM=month and DD=day) and time format (12-hour or
24-hour). Default: YYYY-MM-DD 24 hours.
DATE/TIME
Sets the date and time. Use the arrow buttons to enter the date and time.

Configure the 4-20 mA analog outputs

If an analog output(s) in the analyzer is connected to an external device, select the channel
represented at the analog output and the measurement range.
1. Enable an analog output as follows:
a. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > SETUP OUTPUTS
> 4-20 mA SETUP > [select an output].
b. Select SELECT SOURCE > [analyzer name].
2. Select an option.
Note: Select SET PARAMETER first, then SET FUNCTION and then ACTIVATION.
Option
ACTIVATION
SELECT SOURCE NONE (default)—Sets the analog output to disabled. [analyzer name]—Sets the analog
SET PARAMETER Sets the channel represented at the analog output.
Description
The ACTIVATION options change based on the SET FUNCTION setting. Refer to the
tables that follow to configure the analog output.
output to enabled.
Note: Do not select a channel that contains the symbol "~" (e.g., 4-~SAMPLE4).
Channels that contain the symbol "~" are not measured.
English 13