Verfügbare Sprachen

Performing an assay
in basic measuring mode
In the basic measuring mode the QuikRead go
instrument performs an assay measurement and
shows the result on the display with reagent lot
data.
To perform a measurement do the following:
1. Choose Measure on the main menu and
follow the instructions on the display (see
Picture 18).
2. Drop a cuvette into the measurement well
in the right position. The barcode of the
cuvette must be facing towards you (see
Picture 19).
Note: do not place a finger or any other
object into the measurement well.
3. The lid closes and the instrument starts the
measurement.
4. After the test is completed, the result is
shown on the display and the cuvette lifted
up for removal.
5. Remove the cuvette. The result disappears
from the display. It can be displayed again by
choosing See previous result.
6. If you wish to make another measurement, in-
sert a new cuvette into the measurement well.
Choosing Cancel takes you to the main menu.