Weighing in Percent
Purpose
This application program allows
you to obtain weight readouts
in percent which are in proportion to
a reference weight.
Preparation
Configure the Weighing in Percent
§
application in the operating menu:
See "Configuring the Balance."
Menu code: 2. 1. 5
Example
Determine an unknown percentage: store the weight on the balance as a refer-
ence percentage
Menu: Application program: Weighing in Percent (menu code
Menu: Reference percentage: 100 % (menu code
Step
1. Tare the balance
2. Display the reference
percentage:
3. Place the reference weight
for 100 % on the balance
(here: 222.5 g)
4. Start the application; if the print [FUNCTION]
format is set to include data
ID codes the following is printed
5. Place an unknown weight
on the balance (here: 322.5 g)
6. Display weight
7. Display percentage
8. Unload the balance
9. Delete the reference percentage [CF]
10. Repeat the procedure starting from step 5, if desired.
14
Reference percentage:
$
Code 3. 3. 1
Code 3. 3. 2 10 % (factory setting)
Code 3. 3. 3 20 %
Code 3. 3. 4 50 %
Code 3. 3. 5 100 %
$ Storage parameter
(display accuracy for counting)
Code 3. 4. 1 Standard resolution:
Code 3. 4. 2 With 10 times higher
See also "Configuring the Balance"
3. 3. 5
Key (or instruction)
[TARE]
[FUNCTION]
>2 sec.
[FUNCTION]
[FUNCTION]
5 %
With stability (factory
setting)
stability than standard
)
2. 1. 5
)
Display/Printout
0.0
g
ref100
+
222.5
g
+
100.00
%
Wxx% + 222.500 g
+
144.94
%
+
322.5
g
+
144.94
%
0.00
%