5 .. W hen.the.measuring.cycle.which.lasts.about.20.
seconds. is. completed,. the. measurement. result.
will.be.displayed.as.shown.below:
Rape
11.6% 22.0°C
.
6 .. T he. last. three. results. are. automatically. saved . .
You.can.then.read.the.average.value.of.the.three.
readings.of.measures,.by.pressing.the..
to.provide.the.final.result
Rape
[3] 11.5%
.
. . T he. average. result. is. always. calculated. from.
the. last. 3. measurements . The. number. of.
measurements. taken. is. always. displayed. in.
square. brackets . Results. below. and. above. the.
measuring.scale.are.not.taken.into.consideration
7 .. A fter.each.measurement.you.should.unscrew.the.
chamber.cover.using.a.socket.wrench.(using.the.
rotation. direction. switch). then. empty. and. clean.
the.chamber.together.with.the.cap's.friction.disc.
using.a.special.brush
..button.
8 .. T horough.cleaning.is.particularly.required.when.
changing. the. measured. grain. variety. and. when.
measuring.very.diverse.samples.or.samples.with.
increased.moisture.level
9 .. F riction. elements. must. be. cleaned. immediately.
after. measuring. samples. with. increased.
moisture.level
EN
11