Replace the fuses
Electrocution hazard. Always remove power to the instrument before making electrical connections.
Fire hazard. Use the same type and current rating to replace fuses.
Refer to
Figure 4
and
Fuse specifications:
Relay fuse: T 5.0 A, 250 V
Output power fuse: AC: T 5.0 A, 250 VAC; DC: T 1.6 A, 250 VAC
Input power fuse: AC: T 1.6 A, 250 VAC; DC: T 6.3 A, 250 VAC
Figure 4 Access cover removal
8 English
Figure 5
to replace the fuses.
D A N G E R
D A N G E R