Transport

The sensor unit can be configured in conjunction with the Dräger X-pid xx00
mobile app, see technical manual.
5.7

Replacing the water and dust filter

Replace the water and dust filter if contaminated or blocked.
Only use the water and dust filter listed as a spare part.
Designation and description
Water and dust filter
(without Luer adapter)
Water and dust filter set
(incl. Luer adapter for OD ~5 mm and
3 mm)
5.8

Cleaning

The sensor unit does not require any special care.
● In the case of heavy contamination, the sensor unit can be carefully cleaned
with cold water. A sponge can be used for wiping if necessary.
CAUTION
Damage to the sensor unit
The sensor unit may be damaged by the ingress of liquid.
► Make sure that no liquid penetrates into the gas inlet openings and the gas
outlet opening on the sensor unit.
► Clean carefully as required. Do not use detergents or solvents.
CAUTION
Damage to the activated carbon filter
Detergents and solvents can stress and potentially overstress the internal
activated carbon filter.
► Do not use detergents or solvents near an operating sensor unit.
● Dry the sensor unit with a cloth.
6
The sensor unit contains lithium batteries. During transport of the sensor unit,
and particularly during air transport, observe the relevant safety regulations for
lithium batteries.
Do not transport the sensor unit at below -20 °C or above +50 °C as it could
become damaged.
NOTICE
Order no.
Observe the safety instructions of the control unit!
8319359
The safety instructions of the control unit contain separate information on
storage and must be observed.
8319364
7
Store in a dry environment free of pollutants.
Dräger recommends charging the sensor unit and the control unit during
storage.
Dräger recommends checking the charge status of the energy supply at
least every 3 weeks if the sensor unit or the control unit is not left on charge.
Do not store the sensor unit at below -20 °C or above +50 °C as it could
become damaged.
NOTICE
Observe the safety instructions of the control unit!
The safety instructions of the control unit contain separate information on
storage and must be observed.
WARNING
Risk of explosion!
Discharging isolated capacitances can create ignitable sparks.
► When storing the device, make sure that it is grounded and thus no isolated
capacitance can be electrostatically charged.
To keep power consumption low in the heat-up phase, and to keep the heat-
up phase as short as possible, store the sensor unit at room temperature.
Transport
Storage
Dräger X-pid
®
9000/9500 Sensor Unit