Settings

13.1 Sensor settings

Sensor settings are available in PolySoft.
– Setting the capture range
The capture range allows unwanted zero deviations to be
suppressed. The setting affects the measured value which
is output.
This feature is used to define the capture range and the
capture value limits.
– Setting the calibration interval
The calibration interval defines the time window for the
next calibration. If the interval expires, the gas detector
issues a warning.
– Sensor self-test
The sensor self-test is a safety feature that is performed at
recurring intervals. If the sensor does not pass a sensor
self-test, a warning is output. Further negative self-tests
will result in an error message being output.
The sensor self-test can be disabled in PolySoft or carried
out manually.
For safety-related applications, the sensor self-test must
be activated.
– Gas settings
Depending on the sensor, the gas type, measuring range
and measuring units can be set with this feature.
13.2 Resetting the gas detector to the
factory settings
Can be performed from 10 minutes after turning on the gas
detector.
Device settings, data for ISA100 Wireless
integration, sensor configuration, and Bluetooth
are reset.
The feature is also available in PolySoft.
Procedure using the magnetic pen
Reset device to factory settings:
1. Hold the magnetic pen on the mark (Fig. A 12) for approx.
15 seconds until the blue LED flashes rapidly.
2. Withdraw the magnetic pen.
3. Keep the magnetic pen on the marking for 5 to 10 seconds
(LED stops flashing) and then take it off.
✓ The gas detector restarts and is reset.
13.3 Changeable security-relevant
settings
PolySoft
Factory setting
SIL status
off
Instructions for use
|
Dräger Polytron
network
®
pairing PIN
®
6100 EC WL, Dräger Polytron
PolySoft
Factory setting
000000
®
Bluetooth
pairing PIN
Sample gas
sensor-specific
Sample gas module
sensor-specific
Measuring range
sensor-specific
Upper capture value
sensor-specific
limit
Lower capture value
sensor-specific
limit
Test gas
sensor-specific
Test gas unit
sensor-specific
Test gas concentration
sensor-specific
Calibration interval
sensor-specific
A1 alarm
sensor-specific
A2 alarm
sensor-specific
Capture value
sensor-specific
Sample gas LEL factor
sensor-specific
PROFIsafe Watchdog
60,000
Timeout
PROFIsafe Address F-
1
Host
PROFIsafe Address F-
1
Device
PROFIsafe Data For-
float
mat
ISA 100 Provisioning
Over the Air
ISA 100 Tag Name
Serial number of the device
ISA 100 Network ID
0
14

Transport

The gas detector contains lithium batteries. During transport,
especially when transporting the gas detector by air, observe
the relevant safety regulations and the marking for lithium
batteries.
15

Disposal

This product must not be disposed of as municipal
waste. It is therefore marked with the adjacent symbol.
Dräger will take back this product for free. For informa-
tion please contact the national marketing organiza-
tions or Dräger.
Batteries must not be disposed of as household waste.
They are therefore marked with the adjacent symbol.
Dispose of batteries and rechargeable batteries at bat-
tery collection points and in accordance with current
regulations.
®
Repeater ISA
|
Settings
en
35