11. Specifications
The maximum cable length of all attached cables e.g. temperature probe or interface cables is 2 m [6.56 ft].
To reach the maximal accuracy an ambient temperature between 17 °C and 23 °C [63 °F and 73 °F] must be
fulfilled.
This instrument is intended to be used in a basic electromagnetic environment, e.g. light-industrial locations,
workshops, service centres etc. In the event of interference caused by high-frequency electromagnetic fields in
a frequency range from 380 ... 480 MHz, an increased measuring deviation of up to 0.3 K is expected. To avoid
possible inferences do not install the instrument near to strong radio transmitters.
11.4 Digital indicator
Digital indicator
Display
Screen
Resolution
Display units
Functions
Real-time clock
Voltage supply
Power supply
Permissible ambient conditions
Operating temperature
Relative humidity
Storage temperature
Communication
Interfaces
Case
Dimensions (W x H x D)
Weight
11.5 Certificate
Certificate
4)
Calibration
Recommended recalibration interval
4) System calibration with a temperature probe and/or electrical calibration of the CTR3000 itself
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 60.15 and the order documentation.
WIKA operating instructions, model CTR3000
Colour TFT display including projective capacitive touchscreen with a resolution of 800 x 480 pixels
0.0001 K / 0.00001 Ω / 0.00001 mV
°C, °F, K, mV and Ω
Integrated clock with date
AC 100 ... 240 V, 50/60 Hz, 0.6 A; universal input on rear panel
0 ... 50 °C [32 ... 122 °F]
Maximum achievable accuracy within 17 ... 23 °C [63 ... 73 °F]
0 ... 70 % r. h. (non-condensing)
-20 ... +50 °C [-4 ... +122 °F]
Standard: USB host, USB device and Ethernet
Optional: RS-232
314 x 176 x 322 mm [12.4 x 6.9 x 12.7 in]
6 kg [13.2 lbs]
Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
1 year (dependent on conditions of use)