5.1. Connecting / disconnecting the rotary instrument
To insert and tight the mandrel
- Apply pressure with the thumb to the center of the
push button.
- Put the instrument into position by turning it slightly
in both directions, until it engages, then release the
push button.
- Check that the rotating instrument is properly attached,
through a slight axial movement on each tool change.
Removal of the rotating instrument
- Apply pressure with the thumb to the center of the push button.
- At the same time, remove the rotating instrument.
5.2. Torque selection
- Turn the selector A in the direction of the arrow
while maintaining the B ring fixed.
5
- The torque values are indicated on the instrument.
- The implant manufacturer's recommendations are
to be followed when choosing the torque.
4
4
4
5.3. How to screw and reach the torque
Screwing/ Unscrewing
- Maintain the torque wrench in one hand.
- Turn the cylindrical component B until an audible
click is heard, indicating that declutching has
taken place and that the selected torque has been
attained.
The use of protective gloves is recommended each time the tools are handled. Check the
condition of the rotating instruments to be used, and handle them properly and carefully.
Inspect the push button regularly to ensure that it is properly fixed.