Herunterladen Inhalt Inhalt Diese Seite drucken

Temperature Settings; Operator Control Panel; Safety System - Gaymar Thermacare TC3249 Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 96
Section 8 - Specifications
8.0 Specifications
8.1 Physical
Dimensions
42 cm x 28 cm x 27 cm
Weight
6.8 kg
Enclosure
Thermoplastic
Filter
HEPA filtration down to
0.3 micron particle size
8.2 Electrical
Classification
Type BF, Class 1, grounded
equipment suitable for
continuous operation. Not
classified for protection against
harmful ingress of liquid.
Input
220-240 V (nominal), 50 Hz,
7 amps max
Motor
1/25 HP, single phase
Heater
1200W heating element
Detachable
Use only an International
Cord
(Harmonized) three wire
cordset using cordage
approved to HD-21.
Conductor size is 1.0 mm
(H05VVF3G1.00)
Circuit Breaker 10 amp
Current Leakage (Earth) 150 microamps max.
Ground Resistance 0.15 ohms nominal;
0.50 ohms maximum
Ambient Operating Temperature
16°C to 29°C

8.3 Temperature Settings

The air temperatures are identified on the
operator control panel and indicate the average
air temperature at the hose end using a Quilt.
NOTE: The air temperature around the
patient is affected both by the ambient room
temperature and the use of an insulating
blanket on top of the Quilt.

8.4 Safety System

Dual Patient Safety Temperature Limit
Thermostats
Either of two independently operating
thermostats will shut off the TC3249 at a
preset high limit temperature. The heater
element and blower will remain off until the
thermostat is manually reset.
Heater Overtemp
Heater overtemp thermostat will shut
off the TC3249 in the event of a blower
failure or lack of air movement. The heater
element and blower will remain off until the
2
thermostat is manually reset.
High Temp Indicator
The REMOVE FROM USE indicator will light
on the front panel when a patient safety
thermostat or heater overtemp thermostat
has tripped. Also, the heating element and
blower will turn off.
HEPA Filter Interlock Switch
The REMOVE FROM USE indicator will light
on the front panel when the HEPA filter has
been installed incorrectly or is missing.
Also, the heating element and blower will
turn off.
8.5 Quilts
All Quilts are made of a nonwoven layer bonded
to a plastic film.
Quilt material meets U. S. flammability
standards:
• NFPA 702 "Normal Flammability"
• CPSC Part 1632
• California Title 19, Subchapter 7
Clear drape material meets NFPA 702 "Normal
Flammability. "
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis