standard EN ISO 11607-1 or in any equivalent
standard required by the national regulations.
e) Sterilization
The tips must be sterilized individually in an
autoclave,
according
parameters:
- autoclave: type B, compliant with standard
EN 13060;
- sterilization temperature: 134°C;
- sterilization stabilization time: 18 minutes;
- pressure: 2 Bars min.
f) Storage
Then, store the sterilized products in a dry place
protected from dust.
to
the
following
Before re-use, if the integrity of the packaging is
not certain, repackage then re-sterilize in
accordance with the defined protocol.
If contamination is visible inside the bag, destroy
the product.
g) Disposal of the product
Dispose of the product in receptacles for waste
materials of healthcare activities involving
infectious risks.
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.