6. Commissioning, operation
5. Screw in the instrument by hand, taking care not to cross the threads.
Tighten the pressure transmitter with a suitable torque spanner using the spanner flats.
When there is a cooling element, the lower hexagon should be used for tightening.
The correct torque depends on the dimension of the pressure connection and the sealing used (form/material). The
maximum torque is 50 Nm.
The specified tightening torque for the high-pressure pipes must be adhered to (see pipe supplier's specifications).
Non-compliance can damage the instrument or the measuring point.
For information on tapped holes and welding sockets, see Technical information IN 00.14 at
www.wika.com
WIKA operating instructions pressure transmitter, model IS-3
EN
Spaner flats
37