Resetting to default settings

You can reset the measurement parameters and the configuration to the
delivery status separately from one another (initialization).
The following measurement parameters (pH/O
delivery status:
pH/ORP (pH InI)
Measuring mode
Asymmetry
Slope
Oxygen (O
InI)
2
Measuring mode
Relative slope
Salinity (value)
Salinity (function)
Conductivity (
Measuring mode
Cell constant
The following configuration parameters (InI) can be reset to the delivery
status:
Baud rate
Display during pH calibration
Interval 1
(automatically saved)
Interval 2
(for data transmission)
Reset procedure:
Press <CAL> while pressing the <RUN/ENTER> key.
pH
To reset the measurement parameters, use <▲> <▼> to select YES and
cm
confirm with <RUN/ENTER>, or
Sal
/
K
TP
continue with the next parameters without a reset with the <RUN/
ENTER> key (O
After the InI configuration parameters, the instrument switches to the last
active measurement mode.
pH
0 mV
-59.16 mV
D. O. concentration
1.00
0.0
Off
InI)
0.475 cm
4800
Buffer nominal value
OFF
OFF
InI >
InI > InI).
2
Resetting to default settings
/
InI) can be reset to the
2
-1
27