Herunterladen Inhalt Inhalt Diese Seite drucken

Connector Plug - WIKA Cryo Gauge 712.15.100 Betriebsanleitung

Differenzdruckmessgerät
Vorschau ausblenden Andere Handbücher für Cryo Gauge 712.15.100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

9. Switch contacts
Mounting of switch contacts
GB
First the snap fit style bezel with sealing ring, and the gauge window, must be removed. Prior
to attaching the transparent hood to the measuring instrument, the contacts must be adjusted
in accordance with their operating range.
Using the magnetic snap-action contact, the magnetic retention force must be adapted to the
instrument-specific conditions by adjusting (twisting) the screw-in magnet. The magnet must
then be protected against unintentional adjustment by using an appropriate locking varnish.
The small plate spring at the flexible contact arm must be angled accordingly.
The fully-adjusted unit must now be fitted to the gauge, together with the snap-fit bezel, and
aligned to the extent that the fork coupling guiding the flexible contact arms grips over the
gauge pointer without touching the dial. Should this occur, the carrying fork must be shorte-
ned using a suitable cutting tool.
The contacts are best adjusted when they are factory-fitted.
By snapping the bezel onto the case, the whole contact unit will be fixed to the pressure
gauge.

Connector plug

As counterpart to the connector base welded
onto the transparent hood
Material: PA 6 - GF 30
Colour: grey
Connection: stripped and tinned wires
Ingress protection: IP 65 per EN 60529 / IEC 529
For magnetic snap-action contacts:
Connector plug 3 poles + l (250 V max.)
with 2 m cable 4 x 1.0 mm
For inductive alarm sensors:
Low voltage execution without protective earth conductor
Connector plug 4 poles (50 V max.)
with 2 m cable 4 x 0.75 mm
Extent of delivery: 1 plug with cable,
1 central screw M3 x 20 and 1 gasket
WIKA operating instructions differential pressure gauge models 712.15.100, 732.15.100
14


Legend
 Self-contained unit with switch
contact for level measurement
 Connector plug
2
2
Gasket

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cryo gauge 732.15.100

Inhaltsverzeichnis