DuoBravo is a versatile device, that measures, compares, displays results, adapts itself, gives
current, calculates statistics ... and much more. Nevertheless it is the fundamental target, to be
simple in use for the patient.
For this reason, each kind of therapy is possible without effort in preparation. Beside 4 free
customizable programs in each version, the following number of pre-set programs are available,
that can be used for therapy at once, without any further adjustment.
- DuoBravo U (Urology) = 12 programs
- DuoBravo N (Neurology) = 12 programs
- DuoBravo U + N (Urology - Neurology) = 22 programs

Use of pre-set programs

1. Connect the electrodes to the device (s.p. 64)
2. Place electrodes or probe according to the instructions of your therapist
3. Switch on DuoBravo with the central ON/OFF-button
4. Select the desired program with the PRG-button
5. Press START
- In phases with pure passive stimulation you'll hear a short melody and mA are flashing,
showing that it can be adjusted now with the +button of any used channel. The intensity will
increase 1 mA with each press - please adjust current as far as it is possible without any
pain. 5 seconds after last adjustment stimulation starts. Please note that this increase is only
possible, if the +button is active (s.p. 56 - Safety Adjustments).
- In EMG- or ETS-phases threshold will be always find out automatically by doing two short
work/rest-periods. If that's not intended you can switch off this function by holding the
START-button for three seconds.
- An EMG-phase would start with the first rest-period now. If you are using an ETS-phase the
short melody sounds and flashing mA prompt for intensity-adjustment.
If the selected program has several phases, each phase-end will be indicated by signal beeps.
Please note that stimulation-intensity will fall down to 8 mA each time a new phase starts. The
intensity will need to be adjusted accordingly.
DuoBravo Bedienungsanleitung / Operators Manual
Therapy with DuoBravo
71