Herunterladen Inhalt Inhalt Diese Seite drucken

Protector PHANTOM Gebrauchs- Und Wartungsanleitung Seite 10

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 12
PHANTOM
FITTING SPARES
To Replace the Visor:
1.
Using a Pozi-drive Screwdriver,
unscrew the two visor clamp screws and
remove the visor clamps from the mask.
2.
Remove the visor from the location
groove in the outer mask.
3.
Replace in reverse order, ensuring
that the witness marks on the visor and
frame are in alignment with the centre-
line of the facemask.
4.
Tighten screws evenly. DO NOT
over-tighten.
Note:
If required, the visor clamp grooves may
be lubricated with a solution of soap and
water in order to ease assembly.
To Replace the Exhale Valve:
1.
Holding the outer edge of the
exhale valve cover, pull to remove from
the side fitting.
2.
Grasp the exhale valve flap and pull
gently to remove from the side fitting (see
Fig. 7).
Fig. 7
3.
Position the replacement valve flap
onto the side fitting, ensuring that it lies
flat and is correctly seated.
4.
Replace the exhale valve cover by
aligning it with the side fitting and
applying pressure until the cover snaps
into place.
To Replace the Blower O-Ring:
1.
Withdraw the blower motor from the
facemask, following the instructions given
in this manual under Cleaning.
2.
Remove
blower motor.
3.
Lightly lubricate the replacement O-
Ring with MS4 Silicone Grease.
4.
Fit the replacement O-Ring into the
forward groove of the blower assembly,
taking care to ensure that the O-Ring is
seated correctly.
5.
Re-fit the blower motor to the
facemask.
To Replace the Filter Seal:
1.
Withdraw the blower motor from the
facemask, following the instructions given
in this manual under Cleaning.
2.
Using the O-Ring Tool provided,
gently remove the seal from its location
flange (see Fig. 8).
3.
Position the replacement seal in the
cavity with the moulded number facing
downwards (so that it cannot be seen).
4.
Using the O-Ring Tool, push the
seal until it slips over the flange and
snaps into place. Continue around the
seal until evenly seated.
CAUTION:
Rubber components should always be
stored in clean, dry conditions away
from direct sunlight and heat.
8
the
O-Ring
from
Fig. 8
the

Werbung