6. Commissioning, operation
9. Enter the temperature of your reference now.
10. Change the digits using the arrow key [▲].
11. Use the arrow key [▼] to switch to the next digit.
12. Confirm by pressing [ENTER/MENU] after completing the input of the temperature.
13. Press [ESC] to return to the measuring mode.
EN
The physical calibration oP2 cannot be cancelled by means of the [ESC] key.
If required, the calibration can be cancelled by switching off the measuring
instrument.
Example of the measured value display after physical calibration against a
reference standard:
The small 2 in the left corner in connection with the display segment CAL in the centre of
the display indicates that the physical probe calibration oP2 has been activated.
6.4.4 Activating/deactivating measuring channels [Chnl]
(only for 2-channel instruments)
Chnl = Channel = select
1. Press the [ENTER/MENU] key and select Chnl using the arrow keys [▲] or [▼].
2. Press [ENTER/MENU] key once again.
⇒
A small 1 appears on the left side of the display, it indicates the channel.
3. Use the arrow keys [▲] or [▼] to select the channel to be activated or deactivated.
4. Confirm selection by pressing [ENTER/MENU].
5. Use the arrow keys [▲] or [▼] now to activate on or deactivate off the set channel.
6. Press [ENTER/MENU] to confirm the desired setting.
7. Use [ESC] to return to the measuring mode.
26
WIKA operating instructions, models CTH63I0 and CTH65I0