2.2
PROGRAMMING
The electronics within the 8175 ultrasonic level transmitter allows the unit to be
configurated as either distance, level or volume depending on the application.
To access the CALIBRATION MENU simutaneously press
The 'Enter' key can be locked to avoid accidental or unauthorised access
- section 3.3.
Reference can additionally be made to the menu guide - section 4.2.
2.2.1
PROGRAMMING - DISTANCE
See section 4.4.1
See section 4.4.2
See section 4.4.3
See section 4.4.4
See section 4.4.6
See section 4.4.7
See section 4.4.8
Measurement or Simulation (optional)
QUICKSTART
Choose the
desired language
Select the desired
units
Select the desired
filter
Enter the
Sound velocity
Perform a Teach-In
procedure
Enter the current
output range
Configure relays
1 & 2 (If available)
6
ENTER
Choose filters 0-2 if no echo filtering
of obstructing obstacles is required
and filters 3-9 if echo filtering is
required.
Typically 331.4 m/s within the air
with a deviation of 0.59m/s/°C.
To determine a new zero point and
sound velocity.
For ex. 4mA -> 100cm
and 20mA -> 250cm
See section 2.3
for 5 seconds.
8175