Herunterladen Inhalt Inhalt Diese Seite drucken

Overview Of Important Information - GCE Zen-O RS - 00500 Benutzerhandbuch

Portable oxygen concentrator
Vorschau ausblenden Andere Handbücher für Zen-O RS - 00500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
14. Only use approved accessories with this device. See approved accessories list in section 6.1.
and cannula approved for use with this device. Using unapproved accessories or cannula
may impair the performance of this device.
15. This device is not designed for use with a humidifier or nebuliser. If a humidifier or nebuliser
is used with this device, performance may be diminished and the device may be damaged.
16. Always follow cannula manufacturer's instructions for proper use.
17. Replace the cannula on a regular basis. Check with your home oxygen provider or clinician
to determine how often the cannula should be replaced.
EN
18. Check that this device operates on battery after disconnecting from the power source.
19. Only charge battery in this device or in an approved charger. (See approved accessories
list.)
20. Remove battery if this device is not going to be used for more than seven days. Store bat-
tery in a cool, dry place.
21. Do not use cleaning agents other than those specified in this manual. Allow the cleaning
solution to dry from the cleaned surface before use.
22. Always turn off this device when not in use.
23. Always disconnect power and turn off this device before cleaning. See Chapter 10. Mainte-
nance and Cleaning.
24. Do not obstruct air intake or exhaust vents when operating this device. Blockage can cause
buildup of internal heat and shut down or damage this device.
25. Do not place objects on top of this device.
26. Keep away from children and pets to prevent damage to the device and accessories and/
or inadvertent setting changes.
27. Keep the device away from pets and pests.
28. This device is rated IP22 while used in the carry bag. Do not use in dusty or wet conditions.
29. Always use in a well ventilated location.
30. Always follow the maintenance schedule as specified in Chapter 10.1. Routine Maintenance.
31. If this device indicates an abnormal condition, see Chapter 9. Troubleshooting.
32. Use caution when touching this device in high ambient temperatures.
33. The device can be re-used by a new patient. The device should be cleaned as indicated in
section 10.2 of this user manual and, according to local laws and prescriptions prior to de-
livering to a new patient.

3.3. Overview of Important Information:

1. If an extension cord is necessary, use a UL listed 15 amp or higher cord. Do not connect any
other devices on the same extension cord. Do not use a multisocketed extension cord.
2. Inhale through the nose for the concentrator to work most effectively. Inhaling through the
mouth may result in less effective oxygen therapy.
3. This oxygen concentrator can operate in either continuous flow mode or pulse delivery mode.
Your clinician will provide you with specific instructions for both modes if applicable. See Chap-
ter 14. Technical Description.
6/185

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis