14.1.2. Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Zen-O™ is intended for use in the electromagnetic environment specified below. The customer
or the user of the concentrator should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD)
EN
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions, and volt-
age variations
on power
supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601 test
Compliance
level
± 6kV contact
± 15kV contact
± 8kV air
± 8kV air
± 2kV for power
± 2kV for power
supply lines
supply lines
± 1kV for input/
± 1kV for input/
output lines
output lines
± 1kV differential
± 1kV differential
mode
mode
± 2kV common
± 2kV common
mode
mode
< 5% UT
< 5% UT
(>95% dip in UT)
(>95% dip in UT)
for 0.5 cycle 40%
for 0.5 cycle
UT
40% UT
(60% dip in UT)
(60% dip in UT)
for 5 cycles
for 5 cycles
70% UT
70% UT
(30% dip in UT)
(30% dip in UT)
for 25 cycles
for 25 cycles
<5% UT
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 5 sec
for 5 sec
3 A/m
3 A/m
Electromagnetic environment/
level
Floors should be wood, concrete, or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity should
be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should
be that of a typical commercial or
hospital environment. If the user of
the Zen-O™ Portable Oxygen Concen-
trator required continued operation
during power main interruptions, it is
recommended that the concentrator
be powered from an uninterruptible
power supply or battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or hospital
environment.
20/185