Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the BC 19
The BC 19 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the BC 19 can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the BC 19 as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Separation distance according to frequency of transmitter
80 MHz to 800 MHz d = 1.2
0.12
0.38
1.2
3.8
12
67
(m)
800 MHz to 2.5 GHz d = 2.3
0.23
0.73
2.3
7.3
23

Werbung