Herunterladen Diese Seite drucken

Hach ORBISPHERE 6110 Allgemeines Benutzerhandbuch Seite 13

Package analyzer v3

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
Option
Normalization
Column selection
When all details have been entered, select Ok to confirm.
Other parameters
1. Time and Date - Set the system date and time parameters and the display format. Select Ok to
confirm.
2. System information - Display the system information about the currently configured boards.
Select Exit to exit the screen.
Security and user management
Security management
When the instrument is started for the first time, security is disabled by default. To avoid any
unauthorized access, it is highly recommended that each user be entered into the system and given
appropriate access rights as soon as possible.
1. Configure parameters related to confidentiality. On completion, select Ok to confirm.
Option
Access rights
enabled
Auto logoff
Audit trail
Description
If checked, enter the temperature value for CO
then be made based on the temperature entered rather than the sample
temperature.
From the drop down list, define the three measurements to display on the right side
of the measurement results screen after the measurement process has completed.
Description
When checked (recommended), it is required to log on as a registered user to access
the menus. When disabled (default), all menus are access free and there will be no
name recorded against any actions in the audit trail. To set up valid users, refer to
management
on page 14.
When checked, the user is logged out automatically when the set delay for inactivity is
reached. The instrument goes on standby. Enter a maximum time of inactivity (in
minutes) for all users.
When checked, every action from a user is recorded in an audit file for traceability.
These actions cover the configuration, calibration, and maintenance options. The audit
file is a rolling buffer recording the last 1000 actions.
normalization. CO
calculations will
2
2
User
English 13

Werbung

loading