14
admissible
centering depth 2
mm
NOTE:
Center CFT (50 kN, 70 kN, 120 kN) over diameter D2.
Fig. 5.1
Force transducer installation
Nominal
Centering
(rated) force
diameter
(kN)
5 kN
20 kN
50 kN
70 kN
120 kN
ATTENTION
The compressive force is transferred over the adaptation
pieces flanged on the face (property class at least. 12.9) or the
mounting parts with relevant support surfaces. The length of
the screws must be such that the depth of the tapped blind
hole in the flange of the transducer is utilized. However, the
screws must not rest on the end of the tapped blind hole.
NOTE
When tightening the screwed connections, use an
open‐ended spanner on the transducer housing to hold it
steady (dimension M, Chapter 8).
HBM
Ø D2
M
A
Tightening
H7
torque M
A
D2
-
0.5
-
1
10
2
14
4
21
21
F
M
A
Screws for
(NVm)
transducer
mounting
1 x M 2.5; 12.9
1 x M 4; 12.9
4 x M 4; 12.9
4 x M 5; 12.9
4 x M 8; 12.9
CMC
o
o
45
or 22.5
offset in the drawing
min. floor side
thread reach (mm)
2
3
4
5
8