5.2.2.3. Asymmetrical and Symmetrical biphasic pulsed current (TENS)

• Push button [7] to select . Use the central controller [16] to select the desired channel or to link the channels;
• Push button [8], select
• Pulse width:
Select
with button [9] and adjust its value with the central controller [16];
• Beat frequency:
Select
with button [10] and adjust its value with the central controller [16];
• Make your choice between frequency modulation and burst mode as these two options are mutually exclusive;
• If frequency modulation is desired:
First check burst mode to be off:
• Select
with button [11] and set its value to zero with the central controller [16];
• Modulation frequency (spectrum):
Select
with button [10] and adjust its value with the central controller [16];
• Modulation time (Spectrum):
Select
with button [11] and adjust its value with the central controller [16];
• If burst mode is desired:
First check frequency modulation to be off:
• Select
with button [10] and set its value to zero with the central controller [16];
• Burst frequency:
Select
with button [11] and adjust its value with the central controller [16];
• Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
• Surge program (when desired):
• Ramp up time:
Select
with button [12] and adjust its value with the central controller [16];
• Hold time:
Select
with button [12] and adjust its value with the central controller [16];
• Ramp down time:
Select
with button [12] and adjust its value with the central controller [16];
• Interval time:
Select
with button [12] and adjust its value with the central controller [16];
• Delay time (only available with linked channels):
Push button [13] and select channel 2 with the central controller [16].
Select
with button [12] and adjust its value with the central controller [16];
• Select CC or CV with button [13] and the central controller [16];
• Select the desired channel with button [13] and the central controller [16]. Adjust the intensity with controller [17];
• Push button [14] to start the therapy in case a surge program was selected.
5.2.2.4. Monophasic rectangular –and triangular pulsed current
• Push button [7] to select
channels;
• Push button [8], select
• Pulse width:
Select
with button [9] and adjust its value with the central controller [16];
• Pulse interval:
Select
with button [10] and adjust its value with the central controller [16];
• Treatment time:
Select
with button [12] and adjust its value with the central controller [16];
• Surge program (when desired):
• Ramp up time:
Select
with button [12] and adjust its value with the central controller [16];
• Hold time:
Select
with button [12] and adjust its value with the central controller [16];
• Ramp down time:
Select
with button [12] and adjust its value with the central controller [16];
• Interval time:
Select
with button [12] and adjust its value with the central controller [16];
or
with the central controller [16] and press button [14] to accept;
. Use the central controller [16] to select the desired channel or to link the
or
with the central controller [16] and press button [14] to accept;
37

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis