12.3 Blood pressure (NIBP)
Disinfectable bladderless cuffs: all cuff sizes from the Riester portfolio
Measurement range:
Measurement range
for neonates:
Measurement precision:
Leakage:
Overpressure limit:
Time per measurement
ri-vital
TECHNICAL DATA
76
Systole 25 - 280 mmHg
Diastole 10 - 200 mmHg
MAP 15 - 260 mmHg
Systole 20 - 155 mmHg
Diastole 5 - 110 mmHg
MAP 10 - 130 mmHg
Mean value difference < +/-5mmHg
Standard deviation < 8mmHg
< 6 mmHg /minute.
300 mmHg mode „Adult"
150 mmHg mode „Neo" (neonates)
Overpressure limit (2nd Level): 330 mmHg mode „Adult"
165 mmHg mode „Neo"(neonates)
approx. 25s on average „Adult" mode max. 170 s
„Neo" mode max. 80s
®