STORAGE
The motor should be stored in a clean, dry and vibrations free environment. . If the motor is to be
stored for a long period of time, the machined surfaces protected with anti-rust coating should be
checked and "touched up" if necessary.
If the motor is fitted with anti-condensation heaters, these should be connected during the storage
period.

MOUNTING AND START UP

Electric data of the motor, valid for the rated power, can be found on the nameplate.
It is the responsibility of the user to ensure that the motor nameplate voltage and frequency are
the same as the mains supply.
Connection to the mains supply should only be executed by a recognised professional, respecting
local demands and regulations.
The user must ensure that the diameter of the cable is adequate for the voltage and current
involved. (See nameplate data of motor)
The connection of the wires must be done as described in the connection diagram that is
delivered with the motor. The connection diagram is usually located inside the cover of the
terminal box. If the motor does not have a terminal box, it is located on the housing or attached
*
to the cable.
The connection diagram is structured as follows: (see example)
Brown
Blue
Black
Black
Yellow / green
The leads can be coloured or numbered. Each line represents one lead.
The given connection diagram is an example only.
The motor must be protected against overloads and short-circuits.
When installing the motor, always connect the earth, using the earthing screw in the terminal box
or on the motor frame. Always check local regulations and demands.
The manufacturer cannot be held responsible for damage due to incorrect installation or use.
Free circulation of fresh air for cooling the motor must be guaranteed
The maximum ambient temperature must not exceed the given value that is indicated on the
nameplate and in the motor certificate.
To guarantee the explosion proof protection, certified explosion proof cable glands with sealing
joints adapted to the cable diameters, must be used.
THE MOTOR WILL LOSE ITS EXPLOSION PROOF CHARACTERISTICS IF THE CABLE AND
GLANDS ARE NOT FITTED CORRECTLY!
14/04/2016
Das Dokument darf nur vollständig und unverändert verwendet und weitergegeben werden. Es liegt in der Verantwortung des
Anwenders, die Gültigkeit dieses Dokumentes bezüglich seines Produktes sicher zu stellen.
Seite 62 von 72
ELNOR MOTORS NV
Mains supply
Relay control
**
INSTRUCTIONS MANUAL ATEX_2016 04 14.doc
*or see section „Circuit diagram"
** Motor is wired by VACUUBRAND.
2 / 5