7A
COLLAR
COLLAR MUST BE
PLACED AGAINST
TUBING RETAINER
7B

(3) Cassette Removal

Follow this procedure to unload the Cassette.
1. Set Drive to "0" and turn Drive OFF.
2. Push down on latch side of Cassette and pull latch outward to unhook from rod.
3. Pivot Cassette upwards and unhook Hook Leg from rod and remove Cassette.

(4) Bi-Directional Pumping

Fluids can be pumped into and back out of containers by reversing the drive.
WARNING:
Tubing breakage may result in fluid being
sprayed from pump. Use appropriate measures to
protect operator and equipment.

(5) Partial Bank Pumping

The Pump can be operated with either a partial or full bank of
Cassettes.There is an elastic, VITON
each of the two metal rods of the pump (See Figure 7) that holds the
Cas-settes in place on the pump. These O-rings should be adjusted
as follows. If the pump is partially loaded with Cassettes, slide the
O-rings along the rods, up against the front of the outermost Cassette,
to keep it in place (note position A, for example, in Figure 7). If the
pump is fully loaded with Cassettes, slide the O-rings to the front of
the pump as far as possible (note position B in Figure 8).
8
TUBING
RETAINER
TUBING
LINK
SUPPORT
ROD
FIGURE 7 TUBING LINK AND CASSETTE LOADING
®
fluoroelastomer O-ring on
7C
PRESS HERE TO
AID IN LATCHING
AND UNLATCHING
7D
FIGURE 8 O-RING POSITION