Herunterladen Inhalt Inhalt Diese Seite drucken

Molybdenum Lr With Vario Powder Pack - Lovibond MD110 Bedienungsanleitung

Photometer-system
Vorschau ausblenden Andere Handbücher für MD110:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Methods
GB
Mo 1
Zero
Test
Mo 1
Zero
Test
Mo 1
RESULT
86

Molybdenum LR with VARIO Powder Pack

0.03 – 3.0 mg/l Mo
Fill a clean Mixing Cylinder (25-ml) with 20 ml of the water sample.
Add the contents of one VARIO Molybdenum 1 LR F20 Powder
Pack straight from the foil into the water sample (20 ml).
Close the Mixing Cylinder tightly with a stopper and swirl several times
to dissolve the powder.
Use two clean vials (24 mm Ø) and mark one as blank for zeroing.
Fill each vial with 10 ml of pre prepared water sample.
Close the blank tightly with the cap.
Add 0,5 ml of VARIO Molybdenum 2 LR solution to the sample.
Close the vial tightly with the cap and invert several times to mix the
contents.
Wait for a reaction period of 2 minutes.
Place the blank in the sample chamber making sure that the
are aligned.
Press [ZERO/TEST] key.
The method symbol flashes for approx. 8 seconds.
Remove the vial from the sample chamber.
Place the sample in the sample chamber making sure that the
are aligned.
Press [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Molybdenum.
marks
marks