Herunterladen Inhalt Inhalt Diese Seite drucken

Specifications - WIKA TC59-V Betriebsanleitung

V-pad-rohroberflächen-thermoelement
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

2. Safety / 3. Specifications
2.4 Labelling, safety marks
Product label
EN
Thermocouple in accordance with standard
Explanation of symbols

3. Specifications

Thermcouple types
Type
Recommended max. operating temperature
IEC 60584-1
K
1,200 °C (2,192 °F)
J
750 °C (1,382 °F)
E
900 °C (1,652 °F)
N
1,200 °C (2,192 °F)
Tolerance value
A cold junction temperature of 0 °C (32 °F) is taken as basis with the definition of the
tolerance value of thermocouples.
When using a compensating cable or thermocouple cable, an additional measuring error
must be considered.
8
TC59-V
1x Type K / Standard
Made in Germany 2015
Before mounting and commissioning the instrument, ensure you read
the operating instructions!
ASTM E230
1,260 °C (2,300 °F)
760 °C (1,400 °F)
870 °C (1,598 °F)
1,260 °C (2,300 °F)
Model
11012345
ASTM E230
0 ... +1,260 °C
Year of manufacture
WIKA operating instructions model TC59-V (V-PAD
)
®

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis