Instructions for Use
Table 3:Guidelines and manufacturer's declaration – Electromagnetic
The systems discussed in this manual are intended for use in the type of electromagnetic
environment specified below. The customer or user of the device or system should insure that it is
used in such an environment.
Interference
IEC 60601
immunity tests
test level
Conducted RF
3 V
interference as
150 kHz to
per
80 MHz
IEC 61000-4-6
Radiated RF
3 V/m
interference as
80 MHz to
per
2.5 GHz
IEC 61000-4-3
6 - English
fault tolerance
Conforman
ce level
3 V
rms
3 V/m
3 V/m
Electromagnetic environment - guidelines
Portable and mobile wireless equipment should
not be used at a distance to the device or to the
system that is less than the recommended safe
distance. This is calculated according to the
equation dependent on the broadcast frequency.
Recommended safe distance:
3.5
d
P
V
1
3.5
for 80 MHz to 800 MHz
d
P
E
1
7
for 800 MHz to 2.5 GHz
d
P
E
1
with P as the maximum nominal power output of
the sender in watt (W) according to specifications
of the manufacturer of the sender and d as the
recommended safe distance in meters (m).
The field strength of stationary radio transmitters
should be less than the conformance level
frequencies in accordance with on-site
In the vicinity of devices that bear following
symbol interferences are possible.
A26361-K1001-Z290-2-8N19, edition 1
b
for all
a
testing.