Looking after your Spirometer

Please observe the following precautions:
 Do not touch the screen with fingers. Use only the stylus
provided.
 Use only a damp, lint free, cloth to clean the screen.
 Do not keep the spirometer in a damp place or expose it to
extremes of temperature.
 Do not direct the transducer holder towards a strong light
source whilst operating the spirometer.
 Check the AC charger for compatibility with local power rating.

Product Lifetime

The MicroLoop spirometer is designed for a product lifetime of 7
years.

Disinfection with Perasafe

Step 1: Fill a container with the appropriate volume of lukewarm tap
water. (30-35°)
Step 2: Prepare the solution by adding 81 grams of Perasafe powder
to every 5 litres of water.
Step 3: Stir until the powder dissolves to form a blue solution. Then
leave for 10 – 15 minutes to allow the disinfectant to fully activate.
The lifespan of the Perasafe solution is maximum 24 hours from
activation and/or 20 immersions, whichever is the least.
Immerse the contaminated Transducer entirely in the prepared
Perasafe solution. Follow chapter "Cleaning the Transducer".
Follow the instructions of the chapter "External Surfaces of the
Spirometer" for surface disinfection.
79