4.4. MECHANICAL SPECIFICATIONS
Drop height
Protection from shocks
Vibrations
Watertightness
Dimensions
Weight of set
4.5. ELECTRICAL SPECIFICATIONS
Permanent overload
See graph of maximum permitted peak current on permanent overload, in appendices,
page 50.
The K1 and K2 probes conform to IEC 1010-2-031 / 300 V / Installation category II.
4.6. ELECTROMAGNETIC COMPATIBILITY
Immunity
Generic standard
Fundamental standard : IEC 1000-4-3, aptitude criteria A
Emissivity :
Generic standard
: 1 m on oak receptacle on concrete to IEC 1010
: 100 g to IEC 68-2-27
: 5...15 Hz : 1.5 mm peak
15...25 Hz : 1 mm peak
25...55 Hz : 0.25 mm peak
sweep speed 1 octave/min, duration 10 min
for each axis (3 axes) to IEC 68-2-6
: IP 40 to IEC 529
: probe (W x H x D): 111 x 15 x 25 mm
case (W x H x D) : 135 x 64 x 28 mm
cable attached probe/case : 1.5 m
: 250 g
: EN 50082.1
with maximum degradation of :
K1 probe : - DC current : 15 mV at zero
- AC current (60 Hz) : 2 dB from 10 mA...4.5 A
K2 probe : - DC current : 100 mV at zero
- AC current (60 Hz) : 2 dB from 1 mA...450 mA
: EN 50081.1 negligible level of emissivity of the probe