Fire hazard. Install a 15 A circuit breaker in the power line. A circuit breaker can be the local power
disconnect, if located in close proximity to the equipment.
Electrocution hazard. Protective Earth Ground (PE) connection is required.
Electrocution hazard. Make sure that there is easy access to the local power disconnect.
Connect the sampler to a battery (8754400) or an AC power supply (8754500US, 8754500EU or
8754500UK). Refer to
Figure 9 Connect the sampler to power
Controller connections
Electrical shock hazard. Externally connected equipment must have an applicable country safety
standard assessment.
Figure 10
shows the electrical connectors on the controller.