11 Assistance in case of minor failures

In case of any disturbance in the programme, switch off the scales and disconnect
them from power supply for a while. Then restart the weighing process.
Assistance:
Disturbance
The weight indication does
not appear.
The weight indication is
constantly changing.
The weighing result is obvi-
ously incorrect.
In case of any other error messages, you must restart the scales. Contact the manu-
facturer if the error message persists.
MBD-BA-e-1710
Possible cause
• The scales are off.
• Improperly inserted or discharged batteries.
• No batteries.
• Draught/movements of air.
• Table/floor vibration.
• The weighing pan is in contact with solids or is in-
correctly installed.
• Electromagnetic fields / static charges (choose an-
other location — turn off the device causing the
disturbance, if possible).
• The weight indication was not cleared.
• High temperature fluctuations.
• The warm-up time was not sufficient.
• Electromagnetic fields / static charges (choose an-
other location — turn off the device that causes the
disturbance, if possible).