7.1. REFERENCE CONDITIONS

Influence quantities
Temperature
Relative humidity
Supply voltage
Frequency range
Capacitance in parallel on resistance
Electric fi eld
Magnetic fi eld

7.2. CHARACTERISTICS PER FUNCTION

7.2.1. VOLTAGE
 Characteristics
Measurement range
Resolution
Accuracy
Frequency range
 Input impedance:
750 k
Measured voltage
Input impedance
 Measurement category: 1000 V CAT III or 2500 V CAT I (transients   2.5 kV)
7.2.2. LEAKAGE CURRENT MEASUREMENT
 Before an insulation measurement:
Measurement
0.000 ...
range DC
0.250 nA
Resolution
1 pA
Accuracy
15% + 10 ct

7. SPECIFICATIONS

Reference values
23 ± 3 °C
45 to 55 % RH
DC and 15.3 to 65 Hz
1,0 ... 99,9 V
100 ... 999 V
0,1 V
1 V
1% +5 ct
DC or 15 ... 500 Hz
 
 
to 3 M
depending on voltage measured
1.0 ... 900 V
901 ... 1800 V
 
 
750 k
1.5 M
0.251 ...
10.00 ...
100.0 ...
9.999 nA
99.99 nA
999.9 nA
10 pA
100 pA
10%
104
9 to 12 V
0 µF
nil
< 40 A/m
1000 ... 2500 V
2 V
1% +1 ct
1801 ... 2700 V
 
2.25 M
1.000 ...
10.00 ...
9.999 µA
99.99 µA
1 nA
10 nA
5%
2501 ... 4000 V
2 V
DC
2701 ... 4000 V
 
3 M
100.0 ...
1 0 0 0 ...
999.9 µA
3000 µA
100 nA
1 µA