Herunterladen Diese Seite drucken

Operation; Warranty - testo 506 Bedienungsanleitung

Differenzdruckmessgerät

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
testo 506
Instructions
Suspension
On Button
Readings
Display
Battery
capacity
Battery
Parameters
compartment
(at rear)
Negative
Positive
pressure
pressure
According to the conformity certificate, this product fulfills the gui-
delines in accordance with 89/336/EEC.
0973.5060/04.03/T/wh/05.04.2004
Basic safety instructions
Please read the following safety instructions with care:
Avoid electrical hazards:
Never make measurements with the instruments on or near live compo-
nents.
Preserve product safety:
Operate the instrument only within the parameters specified in the tech-
nical data.
Handle the instrument suitably and according to its intended purpose.
Open the instrument only when this is expressly described in the
Instructions for maintenance purposes.
Never apply force.
Ensure correct disposal:
Dispose of empty batteries at the provided collection points.
Send the unit directly to us at the end of its life cycle. We will ensure
that it is disposed of in an environmentally friendly manner.
Important information
Notes and special cases in the handling of your unit are indicated by an
exclamation mark.
Due to technical reasons, the display speed of the liquid crystal
displays slow down at temperatures below 0 °C (approx. 2 s at -10 °C,
approx. 6 s at -20 °C). Measuring accuracy is not affected.

Warranty

Warranty ........................................................................................2 years
Use of the unit for its intended purpose
The battery-operated testo 506 is used to measure the positive or negati-
ve pressure and differential pressure in gases (See Ordering data for mea-
suring ranges). testo 506 (0 to 500 Pa) is also used, together with a Pitot
tube, to measure air flow.
Its robustness and easy handling make it suitable for customer service and
maintenance work on heating, ventilation and air conditioning units.
All pressures in the positive range can also be measured in the negative
range (Example: In the measuring range 0 to 200 hPa/mbar, the instru-
ment can also measure negative pressure 0 to -200 hPa/mbar). Ensure
connection is correct (positive pressure +, negative pressure -)!
testo 506 zeroises automatically when switched on.
testo 506 switches off automatically after 16 min (Auto Off function).
Battery capacity
Symbol
Capacity
10 - 100%
<10 % (
remaining duration: approx. 15 h)
Changing the battery
It is necessary to change the battery if the symbol
appears in the
display.
-Open housing using a screwdriver.
-Remove spent batteries from battery compartment.
-Place new battery (2 off) in the battery compart-
ment. Insert correctly (+/-)!
+
-Check air density and reset if necessary
(at 0 to 500Pa; standard 1.199 kg/m
3
)
+
-Close battery cover.
-testo 506 is ready for use.
Accessories/Spare parts
Pitot tube
........0635.2245
(300 mm long, Ø 4 mm, stainless steel max. 600 °C)
Pitot tube
......0635.2145
(350 mm long, Ø 7 mm, stainless steel, max. 600 °C)
Pitot tube
(500 mm long, Ø 7 mm, stainless steel, max. 600 °C)
......0635.2045
Pitot tube
....0635.2345
(1000 mm long, Ø 7 mm, stainless steel, max. 600 °C)
Connection hose
(5 m long, silicone, max. 700 hPa)
......................0554.0440
ISO calibration cert.
............0520.0095
(press., 3 points distributed in meas. range)
ISO calibration cert.
(press., 5 points distributed in meas. range)
............0520.0005
Spare battery ..........................................................................0515.0028
Technical data
Common data
Units/Pressure ............................hPa/mbar, mmH2O, mmHg, psi, inH2O
Substances ................................................Air or all non-corrosive gases
Pressure connection ............................................4/6 mm inner diameter
Display ....................................................................................LCD, 1 line
Storage temperature ..........................................................-20 to +70 °C
Operating temperature ............................................................0 to 40 °C
Battery type..................................................2 x 3V button cell (CR 2032)
Battery life ..........................................................................Approx. 150 h
Auto Off..........................................................................................16 min
Dimensions in mm ................................................................86 x 48 x 24
Weight................................................................................................55 g
testo 506 (0 to 10 hPa/4 IWC)
Measuring range ......................................0 to 10 hPa/mbar; 0 to 4 IWC
Accuracy* ........................................................± 0.03 hPa <1 hPa/mbar
......................................................±1.5% of full-scale value >1 hPa/mbar
Resolution ................................................0.01 hPa/mbar; 0.1 mmH2O;
........................................................0.001 inH2O; 0.01 mmHg; 0.001 psi
Overload......................................................................................34 times
testo 506 (0 to 500 Pa/2 IWC)
Measuring range ................................................0 to 500 Pa; 0 to 2 IWC
Accuracy* ..................................................................................± 2.5 Pa
Resolution ............................................1 Pa; 0,1 mmH
O; 0.001 in H
2
................................................................................0.01 mmHg;0.001 psi
Measuring range/Velocity ........................................27 m/s; 5315 ft/min
Resolution/Velocity ......................................................0.01 m/s; 1 ft/min
Overload........................................................................................4 times
testo 506 (0 to 200 hPa/80 IWC)
Measuring range
............................0 to 10 hPa/mbar; 0 to 4 IWC (low)
....................................................10 to 200 hPa/mbar; 4 to 80 IWC (high)
Accuracy* ........................................± 0.03 hPa /mbar (0 to 1 hPa/mbar)
..............................................±1.5% of full-scale value (1 to 10 hPa/mbar)
..................±(1 hPa /mbar+ 1% of measured value) (10 to 200 hPa/mbar)
Resolution
Low: ..0.01 hPa/mbar; 0.1 mmH2O; 0.001 inH2O; 0.01 mmHg; 0.001 psi
High: ............0.1 hPa/mbar; 1 mmH2O; 0.01 inH2O; 0.1 mmHg; 0.01 psi
Overload ....................................................................................1.5 times
* at rated temperature of +22 °C
Measuring
If there is a change in temperature, the instrument
needs 30 minutes to adapt to the new ambient tem-
perature.
- Connect hoses to testo 506.
<1 s
- Switch on testo 506 using On button.
- Segment test is carried out.
Note
Not all segments are available in all of the instrument
versions.
- Automatic initialisation is carried out.
- Measuring mode
- Press On button again to zeroise.
- Connect hoses to pressure system
- Measure
>5 s
- Press On button > 5 s and instrument switches off.
Measuring range has been exceeded
Note
Pressure which is greater than the specified overload
will destroy the sensors.
Switching units
- Keep On button pressed when switching on.
- The unit last set flashes in the display.
- Let go of On button.
- Keep pressing On until the required unit appears in
the display.
O;
2
- Let go of On button.
- Instrument will start with new unit.

Operation

Switching measuring ranges in testo 506
testo 506 (0 to 10/200 hPa/mbar) has an automatic
measuring range switchover function.
The instrument starts in the top measuring range and
switches automatically to the lower measuring range
after 2 s.
If a pressure of 10 hPa/mbar is exceeded, the instru-
ment changes automatically to the higher range
(resolution: 0.01 hPa/mbar
Measuring air flow using testo 506
(0 to 500 Pa) and a Pitot tube
- Keep On button pressed when switching on the
instrument.
- The unit last set flashes in the display.
- Let go of On button.
- Keep pressing On until the m/s (ft/min) unit appears.
Operation
- The value 1.199 for rho in kg/m
shown in the display.
- 1st position flashes
- Keep pressing On until required digit is reached.
- Wait 2 s; value is set;
2 s
cursor changes to the next position.
- 2nd position flashes.
- Keep On button pressed until required digit is rea-
ched.
- Wait 2 s; value is set;
2 s
cursor changes to next position.
etc.
- Continue to do the same for positions 3 and 4
NOTE
This value can be selected between 0.400 (min value)
and 1.599 (max. value).
The set values are retained when testo 506 (0 to 500
Pa) is switched on or off.
Operation
0.1 hPa/mbar).
3
(factory setting) is

Werbung

loading