H61.0.3X.6C-01
4 Handling
4.1 Display
4.2 Basic Operation
ON
OFF
max
1
2
Set
min
Menu
4
5
Please Note: Activating/deactivating tara clears the max- & min-memories.
1)
Zero-Point Adjustment: If there is no pressure or zero-pressure (absolute) applied to the pressure ports the
Please note: - A zero-point adjustment can only be carried out if the difference between the value on display
is less than 500 digits!
- If a zero point adjustment was carried out the display shows "Corr" after a restart .
User's Manual GMH 3181-...
+
Tara
3
ON
Tara
OFF
max
1
2
3
Store
Set
min
Store
Quit
Menu
Quit
6
4
5
6
device will display 0. If there is a permanent deviation (and device is operated
under steady conditions), a permanent zero point adjustment can be carried out. To
carry out the adjustment press button 3 for approx. 5 seconds (Auto Null will be
displayed shortly). The adjustment is done via the OFFSET-value of the sensor
(referring configuration menu).
To recall the manufacturer's calibration press button 3 for approx. 15 seconds.
1
Main display: shows actual value
2
Arrow points to the chosen measuring unit
Secondary display: shows min./max. or hold value
3
SL: appears if sea-level-correction is activated
4
(only GMH 3181-12)
Tara: appears if tara-function is activated
5
Logg: appears if logger function is chosen, flashes while
6
logger is running
On / Off
min/max bei Messung:
press short:
press again:
press 2 sec.:
Tara, zero-point adjustment:
press short:
press 2 sec.:
press 5 sec.:
Set/Menu:
press short:
Store/Quit:
press short:
press again:
at active logger:
shows the min./max. value
hides min./max. value
clears particular value
display will be set to 0
The following measuring will be
relatively displayed to the set tara
value
deactivates tara-function
Zero-Point Adjustment
invokes configuration menu
hold-function, the last measuring
value will be held in the secondary
display.
hides the value
invokes logger functions
page 5 of 16
1)