10. Troubleshooting

Problem/
Possible Cause/Remedy
Question
The atomiser
1. Too much or too little medicine in
produces
the atomiser.
little or no
Minimum: 0.5 ml, Maximum: 8 ml.
aerosol.
2. Check that the nozzles are not
blocked If necessary, clean the
nozzle (e.g. by rinsing). Use the
atomiser again.
CAUTION: Only poke implements
through the fine holes from the
underside of the nozzle.
3. The atomiser is not held in an
upright position.
4. The medicine solution is unsuit-
able for nebulising (e.g. too thick).
The medicine solution should be
specified by the doctor.
Output is too
Hose kinked, filter blocked, too
low.
much inhalation solution.
Which
Consult your doctor.
medicines are
Essentially all medicines that are
suitable for
suitable and approved for device-
inhalation?
based inhalation can be inhaled.
Some inhala-
This is normal and occurs for techni-
tion solution
cal reasons. Stop inhalation if you
remains in the
hear a notable difference in the noise
atomiser.
of the atomiser.
What special
1. In babies, the mask should cover
steps should
the mouth and nose to guarantee
be taken for
effective inhalation.
babies and
2. In children, the mask should also
children?
cover the mouth and nose. Nebu-
lisation next to a sleeping person
is not suitable because insufficient
medicine reaches the lungs.
Note: Children should only use the
device with help and under supervi-
sion of an adult. Never leave a child
alone with the nebuliser.
Inhalation with
This is for technical reasons. Less
the mask takes
medicine is inhaled per breath
longer.
through the mask holes than using
the mouthpiece. The aerosol is
mixed with ambient air through the
holes in the mask.
Problem/
Possible Cause/Remedy
Question
Why do I need
There are two reasons for this:
to replace the
1. To guarantee a therapeutically
atomiser at
regular inter-
vals?
2. You are also recommended to
Does each per-
This is absolutely necessary for
son need their
hygiene reasons.
own atomiser?

11. Technical specifications

Dimensions
(WxHxD)
300 x 180 x 100 mm
Weight
1.9 kg
Operating
pressure
approx. 1.2 bar
Atomiser fill
volume
max. 8 ml
Medicine flow
> 0.2 ml/min.
Sound pressure
< 60 dBA
(acc. to DIN EN 13544-1 section 26)
Mains
connection
230 V ~, 50 Hz; 180 VA
Housing material ABS
Operating
Temperature: +10 °C – +40 °C
conditions
Relative humidity: 10% – 95%
Storage and
Temperature: -20 °C – +70 °C
transport
Relative humidity: 10% – 95%
conditions
Atmospheric pressure: 500 – 1060 hPa
Subject to technical modifications.
Particle size diagram
All measurements were obtained with a sodium chlo-
ride solution using a laser diffraction method.
17
effective particle spectrum, the
nozzle holes should not exceed a
certain diameter. Due to the ther-
mal and mechanical stresses, the
plastic is subjected to a certain
amount of wear. The atomiser [9] is
particularly sensitive. This can also
cause changes to the droplet com-
position of the aerosol, which has a
direct effect on the effectiveness of
the treatment.
replace the atomiser on a regular
basis for hygiene reasons.