Dosing
and
The Auto-Zero function is used to tare small variations in weight
automatically.
Zero-tracking
In the event that small quantities are removed or added to the
material to be weighed, incorrect weighing results can be
displayed due to the "stability compensation" in the balance.
(Example: Slowly draining fluids from a container on the balance).
When apportioning involves small variations of weight, it is
advisable to switch off this function.
If Zero-Tracking however is switched off, the weighing display
becomes more busy.
In weighing mode keep the PRINT key pressed until [Unit]
Press the MODE button several times until „tr" is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key.
Use the SET key to confirm selection.
Selection
In the model series KERN FKB_A, the adjustment weight can be
selected from three pre-set nominal values (approx.1/3; 2/3; max)
adjustment
(refer also to following table, factory setting with grey
weight
background). In order to achieve high-quality weighing results in
the sense of the measuring technology, it is recommended to
select the nominal value as high as possible. The non delivered
adjustment weights can be purchased from KERN as option.
In weighing mode keep the PRINT key pressed until [Unit]
Press the MODE key several times until „CAL" is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key.
Use the SET key to confirm selection.
PCB-BA-e-1718
appears.
Function activated
tr on
Function deactivated
tr off
appears.
29