Guidance And Manufacturer's Declaration-Electromagnetic Immunity - Philips PRP6100/02 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

guidance and Manufacturer's declaration-electromagnetic immunity

The Pain Relief Patch BlueTouch is intended for use in the electromagnetic environment specified
below. The customer or the user of the Pain Relief Patch BlueTouch should assure that it is used in
the following environment:
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE Un is the a.c. mains voltage prior to application of the test level
IEC 60601-1-2
Test Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
<5% Un
(>95% drop in Un)
for 0,5 cycle
40% Un
(60% drop in Un)
for 5 cycles
70% Un
(30% drop in Un)
for 25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
3 A/m
Compliance Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
<5% Un
(>95% drop in Un)
for 0.5 cycle
40% Un
(60% drop in Un)
for 5 cycles
70% Un
(30% drop in Un)
for 25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
3 A/m
Electromagnetic Environment
- guidance
Floors should be made of
wood, concrete, or ceramic
tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the Pain Relief Patch
BlueTouch requires continued
operation during power mains
interruptions, it is recommended
that the Pain Relief Patch
BlueTouch be powered from
an uninterruptible power supply
or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.