Herunterladen Inhalt Inhalt Diese Seite drucken

Burmeier DALI II 24 Volt Gebrauchsanweisung Seite 40

Pflegebett
Vorschau ausblenden Andere Handbücher für DALI II 24 Volt:
Inhaltsverzeichnis

Werbung

Customer / med. Facility / practice:
Address:
Carried out:
Repeat inspection
_____________________________________________
Equipment type:
Hospital bed
Bed type: DALI II 24 Volt
Location:
Grade level (IEC60601-2-52):
1
2
Manufacturer: Burmeier GmbH & Co. KG
Testing equipment used (Type / Inventory Number):
MPG Classification: Class I
I. Visual Inspection
Visual inspection of the electriacl components
Stickers and type plates
Transformer unit housing
Motor housing and lifting rods of motors
Hand switch
Cable for motor and hand control
Plug of harness / switching power supply
Visual inspection of the mechanical components
Stickers and type plates
Safe working load / patient weight
Trapeze pole / socket
Bed frame
Spring wood slats
Castors
Reclining surface
Wooden boards
Welded seams
Side rail
Knurled screws
Wearing parts such as joints
II. Electric measurments according to DIN EN 62353:2008-08 (leakage current, direct)
Proceed as follow:
1.
Plug in transformer unit in test socket of testing device
2.
Insert the plug of the 24 volt cable into the special adapter
(extra accessory; request from BURMEIER)
3.
Connect the special adapter with the test probe of the test unit.
4.
Start procedures with the measuring instrument
Result: Bed SK II (Type B)
40
Nursing bed
3
4
5
Presence, legible
Damage
Correct position, Damage
Damage, Foil intact
Damage, Routing
Available, Correct position
Presence, legible
Presence, legible
Damage, Deformation
Damage, Deformation
Damage, Splinters
Damage
Damage, Deformation
Damage, Splinters
Split welded seams
Damage, Splinters
Tight fit
Damage
Value of first
Limit Value
measurement
0,1 mA
mA
G
EBRAUCHSANWEISUNG
Inspection prior initial operation
Inspection following repairs / servicing
Protection classification:
Inventory Number:
Serial Number:
User-specific: none
1.
2.
Not
OK
Defect Description
OK
Not
OK
Defect Description
OK
Current actual
value
mA
P
DALI II 24 V
· I
FLEGEBETT
OLT
NSTRUCTION
Test protocol of
an inspection of
electro-medical
I
II
M
N
B
ANUAL FOR
URSING
ED
equipment
according to
DIN EN 62353:
2008-08
Page 1 of 2
DALI II 24 V
OLT

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis