Trouble-shooting table aid
Problem
hand switch / drive
system not functioning
Green led on power ad-
apter not lit and drive
systems are without
function
Green led on power
adapter lit continuously,
yellow led not lit when
any hand switch button is
pushed
drives operate in one
direction only
drives stop suddenly
after prolonged
adjustment times
Mattress base not
horizontal
arMiNia iii 24 V
· i
olT
NSTrUcTioN
Possible causes
n hand switch plug not properly
connected
n Transformer plug (24 volt) not
properly connected
n hand switch or drive system
defective
n functions are blocked on
hand switch
n Power cable not properly
connected
n wall socket without power
n Power adapter defective
n hand switch is defective
n Transformer plug (24 volt) not
properly connected
n functions are locked on hand
switch
n hand switch or drive system is
defective
n heat switch in power adapter
has been activated
n lifting actuators not running
synchronously
M
N
b
arMiNia iii 24 V
aNUal for
UrSiNG
ed
Solution
Test the plug connection
Plug the transformer plug into
the connection socket
inform your operator about
necessary repairs
Unlock functions (see
page 25)
Plug in the power cable
properly
Test the wall socket and fuse
box
exchange the power adapter;
inform your operator about
necessary repairs
exchange the hand switch
Test the plug connection
Unlock the functions
(see page 25)
inform your operator about
necessary repairs
Too short minimum pause
(see page 23)
exchange the power adapter;
inform your operator about
necessary repairs
raise or lower mattress base
to highest or lowest setting at
least once a day.