Fault / Display (Printer text)
ERROR [XXX]
..MAIN-SYSTEM
..IR-SYSTEM
..EC-SYSTEM
..FLOW-SYSTEM
..PRESSURE-SYS
..HEATER-SYSTEM
(Message is printed out additionally)
Example:
ERROR [075] HEATER-SYSTEM
Cause
Instrument defective
(XXX = number specifying the error)
Instrument blocks automatically all
measurement sequences when error is
detected.
The predetermined temperature of the
sensor or the hose is not reached.
Fault, Cause and Remedy
Remedy
Switch off instrument.
Check whether the error message
reappears after switching on again. If
yes: inform DrägerService. Bring
instrument to repair.
Ensure compliance with the ambient
conditions. Switch instrument off,
eventually close the instrument cover
and switch on again.
If the error message reappears after
switching on, inform DrägerService.
23