6 Commissioning, Operation
c) Clear manual recordings:
If data sets have been stored, they can be deleted with the button "STORE":
Press for 2 seconds: Call menu "Clear"
Select with
or
Clear nothing (cancel menu)
Clear all recordings
Clear the last recording
Confirm selection and quit menu "Clear"
6.5.2 Automatic Recording with Selectable Cycle Time ("Func CYCL")
If logger function "Func CYCL" is selected (see "Configuration" ) the device will automatically
record measuring values at intervals of the set cycle time.
The logger's cycle time can be set from 1 s to 60 min (see "Configuration").
1-channel version:
Storable data sets: 10000
A data set contains at slow measuring rate (rAtE SLo):
Measuring value at the time of the recording
■
"CALC off":
■
"CALC aktiv"
■
At data set contains at fast measuring rate (rAtE FASt and rAtE P.dEt):
Arithmetic average since the last recording
■
"CALC off":
■
"CALC aktiv"
■
1)
CALC-Function: see chapter 6.4.5
WIKA Operating Instruction, Model CPH6300
min-peak, max-peak since the last recording
1)
: Calculated value at the time of the recording and dummy value
min-peak, max-peak since the last recording
1)
: arithmetic average of calculated value since the last recording
and dummy