Herunterladen Inhalt Inhalt Diese Seite drucken

Valve Manifold With Working Pressure Gauge (Optional) - WIKA 732.15.160 Betriebsanleitung

Differenzdruckmessgerät
Vorschau ausblenden Andere Handbücher für 732.15.160:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5. Valve manifold with working pressure gauge (optional)

5. Valve manifold with working pressure gauge (optional)
The compact optional flanged valve manifold for an NG 100 working pressure gauge allows
the central measurement of level and working pressure in a single device.
GB
Test connection
M20 x 1.5
Isolating valve
i
c) Connect the pressure standard and test pump using the additional G ¼ female port in the
plus chamber of the measuring system
d) Remove the test connection screw from the minus side valve manifold
e) The plus side can then be pressurised
f) After adjustment:
- Close the air bleed screw
- Disconnect the pressure standard and test pump and close the connection
- Slowly open first the plus and then the minus shut-off valve
Test connection M20 x 1.5 for checking the working pressure gauge
The pressure equalising valve allows a zero point control during operation (with open valve).
a) Close the plus and minus shut-off valve
b) Afterwards open the pressure equalising valve
While the media is flowing from the higher pressure side to the other side, the differential
pressure at the pressure gauge drops to zero (the differential pressure display must be at
zero, i.e. within the zero tolerance range which shows that the gauge is working correctly).
A zero adjustment can be made using the standard integra-
ted, adjustable pointer (remove snap fit bezel incl. window and
sealing ring beforehand). Twisting the slotted screw on the
adjustable pointer you can adjust the zero point.
After completing the zero adjustment, the snap fit bezel,
incl. window and ‚o'-ring seal, must be correctly re-fitted
and the pressure equalising valve must then be closed again.
Subsequently the zero point of versions with integrated
transmitter (see page 9) should be checked.
c) Close the pressure equalising valve
d) Slowly open first the plus and then the minus shut-off valve
WIKA operating instructions differential pressure gauge models 712.15.160, 732.15.160
8
To isolate line pressures without inter-
rupting the process, enabling gauge
removal/examination and protecting the
gauge against overpressure of n-times
rated pressure which may occur during
plant pressure testing.
To protect the gauge against pressure
surges/pressure spikes, and thus against
unspecified operating conditions
For gauge shut-downs, if no measure-
ments are required for long operating
periods, i.e. if only occasional measure-
Working
ments are necessary (to increase service
pressure gauge
life of those differential and working pressu-
NS 100
re gauges with a high frequency of pressure
fluctuations).
Pressure equalising
Recalibration of differential pressure
valve
gauges (tank volume display)
a) Close the plus and minus shut-off valve
Isolating valve
b) Afterwards open the pressure equalising
j
valve, wait for a short period of time and then
close the pressure equalising valve again.
Slotted
screw

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

712.15.160

Inhaltsverzeichnis