Herunterladen Inhalt Inhalt Diese Seite drucken

Resistance (Ω) And Continuity; Frequency (Automatic Ranges); Safety; General Information - Chauvin Arnoux F 65 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für F 65:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 36
5.2.5. Resistance (Ω) and continuity
Range
Resolution
0.1Ω
1kΩ
Max. voltage: 3.3V DC during the measurement.
Overload protection: 660 Vrms
Continuity selection threshold: R < 35 Ω

5.2.6. Frequency (automatic ranges)

Function
Ranges
Resolution
A-Hz
0~100 Hz
0.1Hz
A-Hz
100~1 kHz
1Hz
V-Hz
0~100 Hz
0.1Hz
V-Hz
100~1 kHz
1Hz
Frequency measurement for currents greater than 10 mA RMS.
Frequency measurement for voltages greater than 5 V RMS.

5.2.7. Safety

IEC 61010-1, IEC 61010-2-032 and IEC 61010-2-033:
- Insulation: class III
- Pollution level: 2
- Altitude < 2000 m
- Installation category: CAT III 300V

5.2.8. General information

Digital display unit
4 LCD digits with max. reading of 9,999 points
Overload
If a reading overshoots the range, the
displayed.
Battery Low indicator
is displayed when the voltage delivered by the
battery is less than the operating voltage.
Sampling
2 measurements/s for the digital display, 100ms for the
MAX function.
Degree of protection of the enclosure
IP 30 as per NF EN 60529
Maximum opening of jaws
∅ 28 mm
Dimensions
(L x l x H): 218 x 64 x 30 mm
Weight
280 g (with batteries)
34
English
Accuracy
1%+3
Accuracy
0.5%±2cts
0.5%±2cts
symbol is
F 62 – F 65

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

F 62

Inhaltsverzeichnis