Mechanical installation
Transmitter winding
without rotational speed
measuring system
24
Measuring
zone
Flange plane faces
Fig. 5.1
Bolted rotor connection
1. Prior to installation, clean the plane faces of the trans
ducer flange and the counter flange.
For safe torque transfer, the faces must be clean and
free from grease. Use a piece of cloth or paper
soaked in solvent. When cleaning, make sure that you
do not damage the transmitter winding or the rota
tional speed measuring system.
2. For the bolted rotor connection (see Fig. 5.1), use six
or eight DIN EN ISO 4762 hexagon socket screws of
the property class stated in Tab. 5.1, in a suitable
length (dependent on the connection geometry, see
Tab. 5.1 on page 26).
We recommend DIN EN ISO 4762 socket head cap
screws, blackened, smooth‐headed, permitted size
and shape variance in accordance with DIN
ISO 4759, Part 1, product class A.