R e m o t e C o n t r o l
6.10 Supported SCPI commands and data formats
Command
Selection of channel
(HMP2020: OUTPut3, OUT3 and :NSELect {3} not available)
INSTrument
[:SELect] {OUTPut1 | OUTPut2 | OUTPut3 | OUT1 | OUT2 |OUT3} Selection of one channel
[:SELect]?
:NSELect {1|2|3}
:NSELect?
Setting a voltage
[SOURce:]
VOLTage
[:LEVel]
[:IMMediate]
[:AMPLitude] {
voltage
<
[:AMPLitude]? [MIN | MAX]
STEP
[:INCRement) {
<
[:INCRement]? [DEFault]
Setting a current
[SOURce:]
CURRent
[:LEVel]
[:IMMediate]
[:AMPLitude] {
current
<
[:AMPLitude]? [MIN | MAX]
STEP
[:INCRement) {
<
[:INCRement]? [DEFault]
Combined setting of voltage and current
APPLy {
voltage
| DEF | MIN | MAX} [, {
<
>
APPLy?
Turning an output (On/Off)
OUTPut
[:STATe] {OFF|ON|0|1}
[:STATe]?
:SELect {OFF | ON | 0 | 1}
:GENeral {OFF | ON | 0 | 1}
Setting of the OVP (= Over Voltage Protection)
VOLTage
:PROTection
[:LEVel] {
voltage
| MIN | MAX }
<
>
[:LEVel]? [MIN | MAX]
:TRIPped?
42
Subject to change without notice
I MIN I MAX I UP | DOWN }
>
numeric value
| DEFault }
>
| MIN | MAX | UP | DOWN }
>
numeric value
| DEFault }
>
current
| DEF | MIN | MAX}]
<
>
Description
Query of the channel selection
Numerical selection of one channel
Query of the numerical channel selection
Default voltage value of the selected channel
Incremental increase by the previously defined step
Definition of the step size
Default current value of the selected channel
Incremental increase by the previously defined step
Definition of the step size
Combined setting of voltage and current
Turning On/Off the previous selected channel
Query whether the output is on or off
Selection and deselection of all previous selected channels
simultaneously (from firmware version 2.0)
Switching On/Off all previous selected channels simultaneously
(from firmware version 2.0)
Query of the fuse status