Herunterladen Inhalt Inhalt Diese Seite drucken

Sequence Of Temperature-Controlled Adjustment; Sequence Of Time-Controlled Adjustment; Manual Adjustment With Internal Weight - KERN PLJ 600-2GM Betriebsanleitung

Präzisionswaage
Vorschau ausblenden Andere Handbücher für PLJ 600-2GM:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.6.3 Sequence of temperature-controlled adjustment:

At a change in temperature of 3°C the balance announces
the internal adjustment via the display „CALib"
The balance starts the adjustment process and displays
„CAL-t".
If the adjustment process is finished, the balance will
automatically return into the weighing mode.

6.6.4 Sequence of time-controlled adjustment:

After 3 hours, the automatic adjustment is started. „CALib"
is displayed.
The balance starts the adjustment process and displays
„CAL-C".
If the adjustment process is finished, the balance will
automatically return into the weighing mode.

6.6.5 Manual adjustment with internal weight

In weighing mode press
During the adjustment process „CAL-H" is displayed
If the adjustment process is finished, the balance will
automatically return into the weighing mode.
 Use the
Then the balance returns automatically into weighing mode.
PLJ-GM-BA-e-1614
TARE
. „CALib" appears.
F
button to cancel the adjustment process at any
time.
is briefly displayed.